Specifications
<strong>GENERAL</strong>
Housing Dimensions;230.7mm × 166.9mm × 71.9mm (9.1 in. × 6.6 in. × 2.8 in.)
Weight;1.8kg (3.98 lb.), including lithium-ion battery
Battery Model and Type;600-BAT-L-2 (Li-ion)
Battery Life;Over 8 hours for standard operation
Operating Temperature;-10 °C to 50 °C (14 °F to 122 °F) operating temperature range at 80% relative humidity
Storage Temperature;-20 °C to 50 °C (-4 °F to 122 °F) [with batteries]
Display Type;7-inch TFT LCD (thin-film-transistor liquid crystal display)
Display Dimensions (WxH, Diagonal);154.2mm × 85.9mm, 177.8mm (6.1 in. × 3.4 in., 7.0 in.)
Power (DC-IN Voltage);24 VDC (60 W)
DC Connector;Circular, 2.5mm pin adapter, center-positive
Data Storage;64 GB internal storage
<strong>PULSER</strong>
Pulse Voltage;100, 200, 300, 400V, unipolar excitation
Pulse Damping;50 or 400 Ω
Pulse Repetition Frequency;10 to 2000 Hz, “Single Shot” (always equivalent to measurement rate)
Tunable Square Wave Pulser;20 nanoseconds to 5000 nanoseconds (100 kHz to 25 MHz), user adjustable
Pulse Modes;Pulse-Echo, Dual, Thru Transmission (TTU)
<strong>RECEIVER</strong>
Gain;0 to 110dB, 0.1dB increments
Filters;23 Digital Filters*, Narrowband, Wideband, and Low-Pass
Rectification;Fullwave, Half+, Half-, RF (Unrectified)
A-Scan Bit Depth;24 bits
Stored A-Scan Samples;Minimum 1024 saved A-Scan points @ 24bit/point
Amplitude Range;User selectable 0-110%, 0-400%, or 0-800% full screen height
System Linearity;Horizontal: ± 0.5% FSW
Resolution;0.25% FSH, amplifier accuracy ± 1dB
Reject;0 to 85% FSH in 1% increment positions
Measurement Rate;Equivalent to PRF in all modes (single shot)
Sample Rate;User selectable Standard (200 MS/s), Extended Range (25 MS/s), or High Resolution (400 MS/s)
<strong>CALIBRATION</strong>
Automated Calibration;Velocity, Zero Offset Straight Beam (First Backwall or Echo-to-Echo) Angle Beam (Soundpath or Depth)
Test Modes;Pulse Echo, Dual, or Through Transmission
Units;Millimeters, inches, or microseconds
Range;3.5mm to 24,500mm at 5,900m/s (0.138 in. to 1,000 in. at 0.2320 in./µs)
Velocity;635 m/s to 15240 m/s (0.0250 in./µs to 0.6000 in./µs)
Zero Offset;0 to 750 µs
Display Delay;0 microseconds to 2203 microseconds
Refracted Angle;0° to 85° in 0.1° increments, then jump to 90°
<strong>GATES</strong>
Measurement Gates;2 fully independent flaw gates, 1 interface gate
Gate Start;Variable over entire displayed range
Gate Width;Variable from 0.040 microseconds to end of displayed range
Gate Threshold;Variable from 2 to 100% full screen height in 1% increments
<strong>MEASUREMENTS</strong>
Measurement Display Locations;5 locations available (manual or auto selection)
Gate (A, B);Thickness, Soundpath, Projection, Depth, Amplitude, Time-Of-Flight
Echo-to-Echo;Standard Gate B - Gate A
Other Measurements;Overshoot (dB) value for DGS/AVG, ERS (equivalent reflector size) for DGS/AVG, AWS D1.1/D1.5 A, B, C, D values
Sizing Methods;DAC/TCG, TCG Table, AWS D1.1/D1.5, DGS/AVG
<strong>ENVIRONMENTAL</strong>
IP Rating;Rated IP65
Drop-Tested;Method 516.6 Procedure IV, 26 drops, package for shipment
Shock-Tested;MIL-STD-810H, Method 516.5 Procedure I, 6 cycles each axis, 15G, 11 ms half sine 6 cycles each axis, 15G, 11 ms half sine shock testing
Vibration-Tested;MIL-STD-810H, Method 514.5 Procedure I, Annex C, Figure 6, general exposure, 1 hour each axis vibration test
<strong>INPUT / OUTPUT</strong>
Alarm Output;3 Digital Alarm Outputs, 5V TTL
Analog Output;2 Analog Outputs, 0-1 V or 0-10 V scaling
Encoders;2 axis quadrature encoder inputs
Synchronization (Trigger Sync);Sync Input & Output Signals